RH-Pro9 - HVHA Sterilizer 110V-120V PN: RH-Pro9
RH-Pro9 - HVHA Sterilizer 110V-120V PN: RH-Pro9
Couldn't load pickup availability
A proven advancement in “RapidCycle” sterilization, the RH-Pro9 sterilizer will raise your level of sterilization efficiency in a convenient size! Using Cutting-Edge HVHA (High-Velocity Hot Air) technology, the RH-Pro9 provides the fastest instrument turnaround of any FDA-Cleared sterilization technology. No steam or water means no lengthy drying cycle with a longer instrument life free of corrosion.
For Healthcare, the RH-Pro9 can Eliminate Immediate-Use Steam Sterilization, shortened cycles, work-arounds, and Joint Commission scrutiny!
Features
The RH-Pro9’s small size means it fits in most dental cabinets while still featuring three instrument trays to handle your sterilization needs.
- FDA-Cleared 6 Minute Sterilization Cycles(Unwrapped)
- FDA-Cleared 12 Minute Sterilization Cycles (Wrapped)
- Smaller Cabinet-Ready Size (1163 Cubic Inches)
- 65% More Energy Efficient Than Steam
- Non-Corrosive Sterilizer Environment
- Full 3-Year Warranty
- Waterless Sterilization with No Chemicals
- Convenient Smaller Size — Three Instrument Trays
RH-Pro9 Installation and Operation Video
Specifications
Unit Size (OD) | 19.6″ W x 17.5″ D x 13.3″ H 498 mm x 445 mm x 338 mm |
Chamber Size (ID) | 9.5”W x 15.6”D x 7.85”H, 241mm x 396mm x 199mm |
Weight | 60 lbs (27 Kg) |
Instrument Tray | 7.3”W x 12”D x 0.85”H |
Electrical | 110-120V, 60Hz, 1400 Watts warm up, 300 Watts operating 220-240V, 60Hz, 1400 Watts warm up, 200 Watts operating |
Sterilization Temperature | 375°F (190°C) 320°F (160°C) 300°F (149°C) 280°F (138°C) |
Warranty | Three years (parts and labor) |
Safety |
Meets Requirements of U.S. Electrical Safety Standards. |
Sterilization Cycle Times
Total Sterilization Processing Times 375°F
- Unwrapped Instruments – 14 Minutes
- Handpieces – 16 Minutes
- Wrapped / Pouched Instruments – 21 Minutes
- Wrapped Cassettes – 36 Minutes
Total Sterilization Processing Time Low-Temperature
- 320°F – 36 Minutes
- 300°F – 56 Minutes
- 280°F – 126 Minutes
No Dry Cycle
Improves Instrument Turnaround Time and removes the threat of environmental contamination and instrument corrosion from wet instruments.
No Water
Saves Time & Improves Patient Safety – With no water required, the potential for contamination from improperly cleaned water reservoirs is eliminated.
No Extensive Maintenance
There is no costly and time-consuming maintenance – Only a few minutes a week is needed to ensure the safe and effective operation of an RH-Pro Sterilizer.
Low-Temperature Cycle Options
CPAC has created low-temperature cycles to process devices/instruments that are known or could potentially become degraded by RapidHeat’s standard High-Temperature cycle of 375°F.
- Three temperature cycles: 320°F (160°C), 300°F (149°C), 280°F (138°C)
- Choice of three low-temperature cycles
- Allows user to validate and verify 12-log kill sterilization
- Eliminates heat-sensitive device concerns
- Time/temperature profiles equivalent to steam
RapidCycle Sterilization
Unlike traditional Dry Heat, RapidHeat™ is an advanced thermal sterilization technology circulating high velocity hot air in a sealed chamber at 200 to 300 air exchanges per minute.
- Complete cycle from door closed to door open is 20 minutes for wrapped instruments!
- No drying time required
- Easy to use – push button operation